A microfluidic device for continuous manipulation of biological cells using dielectrophoresis.

Abstract:

:The present study demonstrates the design, simulation, fabrication and testing of a label-free continuous manipulation and separation micro-device of particles/biological cells suspended on medium based on conventional dielectrophoresis. The current dielectrophoretic device uses three planner electrodes to generate non-uniform electric field and induces both p-DEP and n-DEP force simultaneously depending on the dielectric properties of the particles and thus influencing at least two types of particles at a time. Numerical simulations were performed to predict the distribution of non-uniform electric field, DEP force and particle trajectories. The device is fabricated utilizing the advantage of bonding between PDMS and SU8 polymer. The p-DEP particles move away from the center of the streamline, while the n-DEP particles will follow the central streamline along the channel length. Dielectrophoretic effects were initially tested using polystyrene beads followed by manipulation of HeLa cells. In the experiment, it was observed that polystyrene beads in DI water always response as n-DEP up to 1MHz frequency, whereas HeLa cells in PBS medium response as n-DEP up to 400kHz frequency and then it experiences p-DEP up to 1MHz. Further, the microscopic observations of DEP responses of HeLa cells were verified by performing trapping experiment at static condition.

journal_name

Med Eng Phys

authors

Das D,Biswas K,Das S

doi

10.1016/j.medengphy.2013.12.010

subject

Has Abstract

pub_date

2014-06-01 00:00:00

pages

726-31

issue

6

eissn

1350-4533

issn

1873-4030

pii

S1350-4533(13)00276-2

journal_volume

36

pub_type

杂志文章
  • Porcine carotid arterial material property alterations with induced atheroma: an in vivo study.

    abstract:OBJECTIVE:A novel methodology has been developed to evaluate regional alterations in arterial wall material properties with induced atheroma in an animal model. METHODS:Atheromatous lesions (fatty, fibro-fatty, and fibrous) were induced in the carotid arteries of a Yucatan miniswine model by endothelial cell denudatio...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2004.09.014

    authors: Nagaraj A,Kim H,Hamilton AJ,Mun JH,Smulevitz B,Kane BJ,Yan LL,Roth SI,McPherson DD,Chandran KB

    更新日期:2005-03-01 00:00:00

  • Development and assessment of a Microsoft Kinect based system for imaging the breast in three dimensions.

    abstract::Three-dimensional surface imaging technologies have been used in the planning and evaluation of breast reconstructive and cosmetic surgery. The aim of this study was to develop a 3D surface imaging system based on the Microsoft Kinect and assess the accuracy and repeatability with which the system could image the brea...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2013.12.018

    authors: Wheat JS,Choppin S,Goyal A

    更新日期:2014-06-01 00:00:00

  • An MRI compatible loading device for the reconstruction of clinically relevant plantar pressure distributions and loading scenarios of the forefoot.

    abstract::The purpose of this study is to demonstrate a new MRI compatible loading device capable of reconstructing realistic loading scenarios of the human foot for research in the field of foot biomechanics. This device has two different configurations: one used to compress the forefoot and one to bend the metatarsophalangeal...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2014.06.006

    authors: Chatzistergos PE,Naemi R,Chockalingam N

    更新日期:2014-09-01 00:00:00

  • Effect of different inertial parameter sets on joint moment calculation during stair ascending and descending.

    abstract::The reliability of internal joint moment calculation in gait analysis during daily living activities is fundamental for clinical decisions based on joint function. This calculation, obtained by means of the inverse dynamics, depends on several modelling factors, such as assumptions on the segments and on the relevant ...

    journal_title:Medical engineering & physics

    pub_type: 临床试验,杂志文章

    doi:10.1016/j.medengphy.2004.12.002

    authors: Fantozzi S,Stagni R,Cappello A,Leardini A

    更新日期:2005-07-01 00:00:00

  • Application of time-varying analysis to diagnostic needle electromyography.

    abstract::Quantification in clinical, diagnostic electromyography (EMG) currently includes motor unit action potential (MUAP) analysis and interference pattern analysis. Early efforts to examine the frequency/power spectra of the interference pattern showed modest value but the technique was not developed further. This paper re...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2011.07.017

    authors: Sheean GL

    更新日期:2012-03-01 00:00:00

  • A practical technique for measuring human biofluid conductivity using high gain-frequency characteristics.

    abstract::Currently, the study of ion composition and performance in human biofluids plays an important role in biomedical engineering research and technology. This field may become universal for human diagnostics; it allows early detection of different diseases in humans by measuring changes in ion behaviour in human biofluids...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2011.04.007

    authors: Al-Nabulsi JI,Aloquili O,Ausheva V,Yuldashev ZM

    更新日期:2011-11-01 00:00:00

  • Liquid ventilation: a mathematical model of gas diffusion in the premature lung.

    abstract::The liquid ventilation (LV) technique was previously demonstrated to be a valuable alternative to ordinary gas ventilation, particularly for newborn patients with severely distressed lungs. This work describes a mathematical model of gas transfer phenomena occurring within the lungs of a preterm newborn baby ventilate...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(96)00041-0

    authors: Costantino ML,Fiore GB

    更新日期:1997-03-01 00:00:00

  • Human body model response to mechanical impulse.

    abstract::Despite our understanding of the importance of falling, previous studies have ignored the effects of falls from standing height. In this article, lumped parameter linear and nonlinear models of a standing human have been used to simulate the impact response of the body segments due to inputs at the shoes. The accelera...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(96)00081-1

    authors: Jarrah M,Qassem W,Othman M,Gdeisat M

    更新日期:1997-06-01 00:00:00

  • A method for determining the heat transfer and water vapour permeability of patient support systems.

    abstract::The formation of pressure ulcers can be exacerbated by a breakdown in the integrity of the patient's skin caused by poor maintenance of the skin microclimate. Patient support systems (PSSs-specialised beds, mattresses, chairs, cushions and pads) play an important role in the dissipation of heat and moisture away from ...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(00)00003-5

    authors: Nicholson GP,Scales JT,Clark RP,de Calcina-Goff ML

    更新日期:1999-12-01 00:00:00

  • A method for the extraction of an object from a noisy background.

    abstract::The removal of the background from a noisy image to leave the object of interest, with well defined edges, is a difficult task. Images obtained using radioisotopes have low signal-to-noise ratios due to scattered photons, as well as emitted radiation from the background. Moreover, noise is, also, a consequence of the ...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/1350-4533(95)00043-7

    authors: Livieratos GN,Chappell PH

    更新日期:1996-03-01 00:00:00

  • New equations to calculate 3D joint centres in the lower extremities.

    abstract::Biomechanical movement analysis in 3D requires estimation of joint centres in the lower extremities and this estimation is based on extrapolation from markers placed on anatomical landmarks. The purpose of the present study was to quantify the accuracy of three established set of equations and provide new improved equ...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2015.07.001

    authors: Sandau M,Heimbürger RV,Villa C,Jensen KE,Moeslund TB,Aanæs H,Alkjær T,Simonsen EB

    更新日期:2015-10-01 00:00:00

  • Quantification of holographic fringe data: comparison of intact and implanted femurs.

    abstract::The strains developed on the surface of a femur with and without an implanted femoral hip replacement were investigated using holographic interferometry (HI). Interferometric fringe data from the four aspects of the loaded femur were quantified and the resulting curvature calculated. Curvature was related to the strai...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(98)00003-4

    authors: Katz DM,Blatcher S,Shelton JC

    更新日期:1998-03-01 00:00:00

  • Suspension microarrays for the identification of the response patterns in hyperinflammatory diseases.

    abstract::Miniaturized and parallelized sandwich immunoassays allow the simultaneous analysis of a variety of parameters in a single experiment. Bead-based protein array systems or suspension microarrays are well-established multiplex sandwich immunoassay formats. To study inflammatory diseases, protein arrays can be used to an...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2008.01.003

    authors: Hsu HY,Wittemann S,Schneider EM,Weiss M,Joos TO

    更新日期:2008-10-01 00:00:00

  • An automated ECG-artifact removal method for trunk muscle surface EMG recordings.

    abstract::This study aimed at developing a method for automated electrocardiography (ECG) artifact detection and removal from trunk electromyography signals. Independent Component Analysis (ICA) method was applied to the simulated data set of ECG-corrupted surface electromyography (SEMG) signals. Independent Components (ICs) co...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2010.05.007

    authors: Mak JN,Hu Y,Luk KD

    更新日期:2010-10-01 00:00:00

  • Reconstruction of proximal humeral fractures with a reduced number of screws and a reinforced bone substitute.

    abstract::Multi-fragmented fractures of the proximal humerus are difficult to treat, especially in the case of osteoporotic bone. Intra-operative risks include cartilage damage when inserting multiple screws. A common post-operative complication is distal-varus collapse of the head. The aim of this study was to investigate if a...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2020.06.002

    authors: Cristofolini L,Morellato K,Cavallo M,Guerra E

    更新日期:2020-08-01 00:00:00

  • Grasping soft tissue by means of vacuum technique.

    abstract:INTRODUCTION:A notable characteristic of bariatric surgery is the frequent manipulation of the bowel. The bowel is large, delicate, flexible, and has a natural lubricant on the tissue surface. Therefore the bowel is difficult to grasp and manipulate. Vacuum technique is commonly used in industry for all types of graspi...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2011.11.014

    authors: Vonck D,Jakimowicz JJ,Lopuhaä HP,Goossens RH

    更新日期:2012-10-01 00:00:00

  • Automated analysis of MR image of hip: geometrical evaluation of the Legg-Calve-Perthes disease.

    abstract::This study proposes semi-automatic determination of geometrical features in hip magnetic resonance (MR) images in order to evaluate the Legg-Calvé-Perthes disease (LCPD). Nine anatomical points on a hip image are selected by a clinician; then eight geometrical indexes of the hip joint are calculated: acetabulum head i...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2004.10.003

    authors: Pouletaut P,Claude I,Winzenrieth R,Ho Ba Tho MC,Sebag G

    更新日期:2005-06-01 00:00:00

  • Stimulated production of vowel-like LX-waveforms and spectral damping in the absence of phonation.

    abstract::Electrical impedance 'LX' waveforms measured across the neck at the thyroid level during phonation are known to be correlated with vocal fold movement. Changes in vocal fold contact are thought to be the cause of this phenomenon though emerging applications in radiotherapy indicate that changes in the configuration of...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(02)00065-6

    authors: Moore C,Jones S

    更新日期:2002-09-01 00:00:00

  • Regression-based indices of fatigue in paraplegics' electrically stimulated quadriceps.

    abstract::We studied muscle fatigue in 19 paraplegic patients during an isometric sustained contraction. Both quadriceps (n = 38) were maximally stimulated for 126 s. We defined a nonlinear equation with four coefficients to fit the data and extract fatigue indices carrying information on the amount (asymptotic values, torque c...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(97)00038-6

    authors: Rabischong E,Chavet P

    更新日期:1997-12-01 00:00:00

  • Shape-memory-alloy-based smart knee spacer for total knee arthroplasty: 3D CAD modelling and a computational study.

    abstract::This study introduced a shape memory alloy (SMA)-based smart knee spacer for total knee arthroplasty (TKA). Subsequently, a 3D CAD model of a smart tibial component of TKA was designed in Solidworks software, and verified using a finite element analysis in ANSYS Workbench. The two major properties of the SMA (NiTi), t...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2018.03.001

    authors: Gautam A,Callejas MA,Acharyya A,Acharyya SG

    更新日期:2018-05-01 00:00:00

  • BION microstimulators: a case study in the engineering of an electronic implantable medical device.

    abstract::The BION (Bionic Neuron) is a single channel implantable neurostimulator of unique design that can be delivered by injection. The development of the BION injectable neurostimulators exemplifies a challenging, but well posed medical design problem addressed with a successful strategy for prioritizing and resolving the ...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章,评审

    doi:10.1016/j.medengphy.2010.08.010

    authors: Kane MJ,Breen PP,Quondamatteo F,ÓLaighin G

    更新日期:2011-01-01 00:00:00

  • A robust transcutaneous electro-muscle stimulator (RTES): a multi-modality tool.

    abstract::This paper introduces a transcutaneous electro-muscle stimulator design that has a wide range of output capabilities. Because of this, the unit is referred to as a robust transcutaneous electro-muscle stimulator (RTES). The RTES is a constant current stimulator that is designed to be capable of generating significant ...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/1350-4533(95)90857-8

    authors: McPartland MD,Mook DJ

    更新日期:1995-06-01 00:00:00

  • On a "Columbus' Egg": Modeling the shape of asymptomatic, dysplastic and impinged hip joints.

    abstract::Understanding morphological features that characterize normal hip joint is critical and necessary for a more comprehensive definition of pathological presentations, such as femoroacetabular impingement and hip dysplasia. Based on anatomical observations that articular surfaces of synovial joints are better represented...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2018.07.001

    authors: Lopes DS,Pires SM,Mascarenhas VV,Silva MT,Jorge JA

    更新日期:2018-09-01 00:00:00

  • Robust infrarenal aortic aneurysm lumen centerline detection for rupture status classification.

    abstract::The objective of this work is to develop a robust method for human abdominal aortic aneurysm (AAA) centerline detection that can contribute to the accurate computation of features for the prediction of AAA rupture risk. A semiautomatic algorithm is proposed for detecting the lumen centerline in contrast-enhanced abdom...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2013.03.005

    authors: Zhang H,Kheyfets VO,Finol EA

    更新日期:2013-09-01 00:00:00

  • Measurement of thermal conductivity of bovine cortical bone.

    abstract::The thermal conductivity of cortical bone was characterized experimentally. Specimens were taken from the mid-diaphysis of bovine femora, and the rate of heat transfer was measured in three orthogonal directions. The conductivity was found to be 0.58+/-0.018 W/mK in the longitudinal direction, 0.53+/-0.030 W/mK in the...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(01)00003-0

    authors: Davidson SR,James DF

    更新日期:2000-12-01 00:00:00

  • Automatic detection of EEG artefacts arising from head movements using EEG and gyroscope signals.

    abstract::Contamination of EEG signals by artefacts arising from head movements has been a serious obstacle in the deployment of automatic neurological event detection systems in ambulatory EEG. In this paper, we present work on categorizing these head-movement artefacts as one distinct class and on using support vector machine...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2012.08.017

    authors: O'Regan S,Faul S,Marnane W

    更新日期:2013-07-01 00:00:00

  • Magnetic resonance-based thermometry during laser ablation on ex-vivo swine pancreas and liver.

    abstract::Laser Ablation (LA) is a minimally-invasive procedure for tumor treatment. LA outcomes depend on the heat distribution inside tissues and require accurate temperature measurement during the procedure. Magnetic resonance imaging (MRI) allows a non-invasive and three-dimensional thermometry of the organ undergoing LA. I...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2015.04.001

    authors: Allegretti G,Saccomandi P,Giurazza F,Caponero MA,Frauenfelder G,Di Matteo FM,Beomonte Zobel B,Silvestri S,Schena E

    更新日期:2015-07-01 00:00:00

  • Fuzzy clustering of gait patterns of patients after ankle arthrodesis based on kinematic parameters.

    abstract::Kinematic parameters for 10 normal subjects and 10 patients with ankle arthrodesis are grouped using the fuzzy cluster paradigm. The features chosen for clustering are Euler angles of the sagittal plane in the hindfoot, the forefoot and combined hindfoot and forefoot joints. Gait patterns are identified using informat...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/s1350-4533(01)00020-0

    authors: Su F,Wu W,Cheng Y,Chou Y

    更新日期:2001-03-01 00:00:00

  • The integrated design of mechanical bi-leaflet prosthetic heart valves.

    abstract::A flow model utilizing an irrotational, inviscid algorithm of vortex-ring elements simulating the leaflets and source/sink elements simulating the aortic root coupled with a boundary layer model has been developed to model the internal flow phenomena of bi-leaflet mechanical heart valves implanted in the aortic root. ...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/1350-4533(95)00075-5

    authors: David T,Hsu CH

    更新日期:1996-09-01 00:00:00

  • Time-dependent analysis of leaflets in mechanical aortic bileaflet heart valves in closing phase using the finite strip method.

    abstract:BACKGROUND AND AIMS OF THE STUDY:Mechanical heart valves (MHV) are widely used to replace dysfunctional and failed heart valves. The bileaflet MHV design is very popular due to its superior hemodynamics. Since their introduction in 1977, the hemodynamics of bileaflet prostheses has been extensively studied. In this stu...

    journal_title:Medical engineering & physics

    pub_type: 杂志文章

    doi:10.1016/j.medengphy.2005.03.013

    authors: Mohammadi H,Ahmadian MT,Wan WK

    更新日期:2006-03-01 00:00:00