Accuracy of the Microlife large-extra large-sized cuff (32-52 cm) coupled to an automatic oscillometric device.

Abstract:

:To determine the accuracy of the large-extra large-sized (L-XL) cuff (32-52 cm) coupled to a Microlife WatchBP Office ABI blood pressure measuring device tested according to the requirements of the International Protocol of the European Society of Hypertension. The L-XL cuff tested in this study is designed to provide accurate blood pressure measurements in patients with large arms (arm circumference≥32 cm) over a wide range of arm circumferences using a single 145±1×320±1 mm bladder. The evaluation was made in 33 patients with a mean±standard deviation age of 53±17 years (range: 30-96 years). Their systolic blood pressure (SBP) was 142±21 mmHg (range: 110-180 mmHg), diastolic blood pressure (DBP) was 87±14 mmHg (range: 62-106 mmHg) and arm circumference was 36±5 cm (range: 32-50 cm). Blood pressure measurements were made in the sitting position. The L-XL cuff coupled to the WatchBP Office ABI passed all three phases of the European Society of Hypertension protocol for SBP and DBP. Mean blood pressure differences between device and observer were -1.3±5.1 mmHg for SBP and -1.8±5.8 mmHg for DBP. Similar device-observer differences were observed in patients divided into two subgroups according to whether their arm circumference was above or below the median in the group. These results indicate that the L-XL cuff coupled to the WatchBP Office ABI monitor provides accurate blood pressure readings in patients with large arms over a wide range of arm circumferences.

journal_name

Blood Press Monit

authors

Masiero S,Saladini F,Benetti E,Palatini P

doi

10.1097/MBP.0b013e328344c73c

subject

Has Abstract

pub_date

2011-04-01 00:00:00

pages

99-102

issue

2

eissn

1359-5237

issn

1473-5725

journal_volume

16

pub_type

杂志文章
  • Reliability of home blood pressure monitoring: in the context of validation and accuracy.

    abstract:OBJECTIVE:Home blood pressure (BP) monitoring offers clinically relevant information enriched with more abundant data. However, there are few studies addressing the reliability of home BP devices and the quality of its data. This study aimed to evaluate the current status of home BP devices in terms of validation and a...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000121

    authors: Jung MH,Kim GH,Kim JH,Moon KW,Yoo KD,Rho TH,Kim CM

    更新日期:2015-08-01 00:00:00

  • Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor.

    abstract:BACKGROUND:To test the accuracy of a new algorithm for the BPM-100, an automated oscillometric blood pressure (BP) monitor, using stored data from an independently conducted validation trial comparing the BPM-100(Beta) with a mercury sphygmomanometer. DESIGN:Raw pulse wave and cuff pressure data were stored electronic...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200106000-00008

    authors: Wright JM,Mattu GS,Perry TL Jr,Gelferc ME,Strange KD,Zorn A,Chen Y

    更新日期:2001-06-01 00:00:00

  • A psychophysiologic reaction to application of ambulatory blood pressure monitor: fact or artifact?

    abstract:BACKGROUND:The study investigated the clinical significance of relatively high blood pressure (BP) elevations during the calibration of the ambulatory blood pressure monitor (ABPM). METHODS:Inferences were drawn from the analysis of differences between 'M', the average of five manually triggered responses, and 'W', th...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e328340ff51

    authors: Yanovski A,Townsend RR,Virginia F,Michael C

    更新日期:2011-02-01 00:00:00

  • White coat hypertension in children.

    abstract::Ambulatory blood pressure monitoring (ABPM) has become more widely used in the assessment of elevated blood pressure in children. The accurate diagnosis of white coat hypertension (WCH) is particularly important in children because detection of elevated blood pressure often results in expensive and invasive diagnostic...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,评审

    doi:10.1097/00126097-200008000-00001

    authors: Sorof JM

    更新日期:2000-08-01 00:00:00

  • Effect of fasting on 24-h blood pressure values of individuals with no previous history of hypertension.

    abstract:OBJECTIVE:We aimed to analyze the difference in 24-h blood pressure values during Ramadan of fasting and nonfasting individuals with no previous history of hypertension. PATIENTS AND METHODS:This study was planned as a multicenter research study in the cities of Izmit, Zonguldak, Sivas, and Adana. The percentage chang...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,多中心研究

    doi:10.1097/MBP.0000000000000244

    authors: Seker A,Demirci H,Ocakoglu G,Aydin U,Ucar H,Yildiz G,Yaman O

    更新日期:2017-10-01 00:00:00

  • Identification of subjects with white-coat hypertension and persistently normal ambulatory blood pressure.

    abstract::OBJECTIVE: To assess the spontaneous changes in clinic blood pressure, ambulatory blood pressure (ABP) and left ventricular structure in untreated subjects with white-coat hypertension (WCH). DESIGN: A prospective observational study. PATIENTS AND METHODS: In 83 untreated subjects with WCH, 24 h non-invasive ABP monit...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Verdecchia P,Schillaci G,Borgioni C,Ciucci A,Gattobigio R,Sacchi N,Guerrieri M,Comparato E,Porcellati C

    更新日期:1996-06-01 00:00:00

  • Comparison of central blood pressure devices on the basis of a modified protocol of the European Society of Hypertension: application to the Centron cBP301.

    abstract:OBJECTIVES:Central blood pressure (cBP) and pulse pressure amplification (PPA) are receiving renewed interest with the increase in the availability of noninvasive techniques that enable its measurement. However, to date, there is no standardized protocol to validate their accuracy. Although invasive comparison seems in...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000028

    authors: Agnoletti D,Millasseau S,Topouchian J,Safar ME,Blacher J

    更新日期:2014-04-01 00:00:00

  • Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure.

    abstract:OBJECTIVES:To determine the accuracy of the A&D UB-511 (UB-512) oscillometric wrist-cuff device for self-measurement of blood pressure, the only difference between the two devices being the size of storage memory. METHODS:Device evaluation was performed according to the modified British Hypertension Society protocol r...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/01.mbp.0000217995.12215.f2

    authors: Angeli F,Sardone M,Angeli E,Repaci S,Gattobigio R,Verdecchia P

    更新日期:2006-12-01 00:00:00

  • Benazepril versus felodipine as supplement to bendroflumethiazide: evaluation by office and ambulatory blood pressure.

    abstract::OBJECTIVE: To compare a combination of a thiazide diuretic and an angiotensin converting enzyme inhibitor with a thiazide diuretic and a calcium antagonist. DESIGN: A double-blind randomized trial with subjects in two parallel groups administered either 10-20 mg benazepril once daily or 5-10 mg extended-release felodi...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Kristensen KS,Wiinberg N,Høegholm A,Kornerup HJ,Svendsen TL,Mølby L,Pindborg T,Nielsen PE

    更新日期:1998-04-01 00:00:00

  • Dialysis-induced hypotension is associated with impaired aortic elasticity in patients undergoing chronic hemodialysis.

    abstract:BACKGROUND:The pathogenesis of dialysis-induced hypotension (DIH) is multifactorial and may include increased aortic stiffness. This study was undertaken to determine aortic elastic properties in patients undergoing hemodialysis with and without DIH, and to assess the effects of DIH on mortality. METHODS:Aortic stiffn...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,多中心研究

    doi:10.1097/MBP.0b013e3282f13f75

    authors: Duman D,Demirtunc R,Erdogan B,Sahin GM,Karadag B

    更新日期:2008-04-01 00:00:00

  • Rationale and design of HOMED-BP Study: hypertension objective treatment based on measurement by electrical devices of blood pressure study.

    abstract:BACKGROUND:How far should blood pressure (BP) be lowered to achieve the greatest reduction in the risk of cardiovascular disease? Although a few trials have tried to answer this question, they failed to convincingly establish an optimal target BP level, in part because of poor reproducibility and the wide variability o...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章,随机对照试验

    doi:10.1097/00126097-200202000-00016

    authors: Fujiwara T,Nishimura T,Ohkuko T,Imai Y,HOMED-BP Study Group.

    更新日期:2002-02-01 00:00:00

  • The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control.

    abstract:OBJECTIVE:The German Hypertension League (Deutsche Hochdruckliga) established a program to assess the accuracy and reliability of blood pressure (BP)-measuring devices in 1999 (Quality Seal Protocol). Here, we report on the results of a testing series of 105 devices designed for BP self-measurement. METHODS:The test p...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000186

    authors: Tholl U,Lüders S,Bramlage P,Dechend R,Eckert S,Mengden T,Nürnberger J,Sanner B,Anlauf M

    更新日期:2016-08-01 00:00:00

  • Effects of the angiotensin II receptor blockers telmisartan versus valsartan in combination with hydrochlorothiazide: a large, confirmatory trial.

    abstract::In 2004-2005, the antihypertensive effects of telmisartan 80 mg versus valsartan 160 mg combined with hydrochlorothiazide 25 mg were assessed in a large placebo-controlled trial in patients with stages 1 and 2 hypertension and demonstrated that both agents were highly effective in lowering blood pressure (BP) compared...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0b013e3282f3859d

    authors: White WB,Murwin D,Chrysant SG,Koval SE,Davidai G,Guthrie R

    更新日期:2008-02-01 00:00:00

  • The effects of renal denervation on resistant hypertension patients: a meta-analysis.

    abstract::We carried out this meta-analysis to assess the effects of renal denervation (RDN) on resistant hypertension patients. According to the collaborative review group search strategy, we searched MEDLINE (1996 to 2015.10); EBCO (1996 to 2015.10) and CNKI. A meta-analysis was carried out using RevMan 5.0. We identified 11 ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,meta分析,评审

    doi:10.1097/MBP.0000000000000177

    authors: Zhang X,Wu N,Yan W,Zhou C,Guo H

    更新日期:2016-08-01 00:00:00

  • Blood pressure assessment during resistance exercise: comparison between auscultation and Finapres.

    abstract:OBJECTIVES:The objective of this study was to test whether the auscultatory method and Finapres provide similar readings for blood pressure response during two resistance exercise protocols of different intensities, and to gain some insight into whether auscultation can be used to monitor cardiovascular stress during r...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e32809ef9f1

    authors: Polito MD,Farinatti PT,Lira VA,Nobrega AC

    更新日期:2007-04-01 00:00:00

  • Determination of accuracies of 10 models of home blood pressure monitors using an oscillometric simulator.

    abstract:BACKGROUND:Patients and doctors often use home blood pressure monitoring (HBPM) to assess the control of hypertension. Despite its popularity there has always been some uncertainty with regard to its accuracy, reliability, reproducibility, and comparability. Although there are pre-market HBPM standards of accuracy, the...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-199904010-00007

    authors: Yarows SA,Amerena JV

    更新日期:1999-02-01 00:00:00

  • Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010.

    abstract:OBJECTIVE:The validation of sphygmomanometer is important in accurate blood pressure measurement. This study presents the validation results by the Medipro MediCare 100f upper arm blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. METHODS:The ESH-IP ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e328346d715

    authors: Yi J,Wan Y,Pan F,Yu X,Zhao H,Shang F,Xu Y

    更新日期:2011-08-01 00:00:00

  • Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypertension Society protocol, the International Protocol and the Association for the Advancement of Medical Instrumentation standards

    abstract:OBJECTIVE:SunTech 247 is an oscillometric blood pressure (BP) monitor manufactured by SunTech Medical Inc., Morrisville, North Carolina, USA. The aims of this study were (i) to determine its accuracy using the 1993 modified British Hypertension Society (BHS) protocol, the 2002 International Protocol (IP) and the 2002/2...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e328330c873

    authors: Polo Friz H,Facchetti R,Primitz L,Beltrame L,Galbiati V,Ricioppo A,Bombelli M,Sega R

    更新日期:2009-10-01 00:00:00

  • Intense aerobic exercise lowers blood pressure in individuals with metabolic syndrome taking antihypertensive medicine.

    abstract:OBJECTIVE:There is a growing tendency for physicians to prescribe exercise in accordance with the 'exercise is medicine' global health initiative. However, the exercise-pharmacologic interactions for controlling blood pressure are not well described. Our purpose was to study whether angiotensin II receptor type 1 block...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000328

    authors: Ramirez-Jimenez M,Fernandez-Elias V,Morales-Palomo F,Ortega JF,Mora-Rodriguez R

    更新日期:2018-10-01 00:00:00

  • Frequency of spontaneous fist clenching during routine blood pressure measurements and its effect on measurement accuracy.

    abstract:BACKGROUND:Guidelines for office blood pressure reading techniques do not warn us about the possibility that patients may clench their fist during cuff-inflation. It is unknown how often patients do this, and what effect it has on measured blood pressure readings. DESIGN AND METHODS:We registered double blood pressure...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200206000-00001

    authors: Smulders YM,Godfried MH,van Montfrans GA

    更新日期:2002-06-01 00:00:00

  • High-intensity handgrip training lowers blood pressure and increases heart rate complexity among postmenopausal women: a pilot study.

    abstract:OBJECTIVE:Handgrip exercise is an emerging strategy for resting blood pressure (BP) reduction requiring minimal time and exercise effort. However, the research literature is currently limited to handgrip protocol designs predominantly prescribing sustained grip contractions, with little assessment of alternative option...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000313

    authors: Bentley DC,Nguyen CHP,Thomas SG

    更新日期:2018-04-01 00:00:00

  • Feasibility of oscillometric aortic pressure and stiffness assessment using the VaSera VS-1500: comparison with a common tonometric method.

    abstract:OBJECTIVES:A number of operator-independent oscillometric devices to measure hemodynamics and arterial stiffness became available recently, but some and in particular VaSera VS-1500 do not provide estimates of aortic pressures and aortic pulse wave velocity (aPWV). The aim of this work was the retrospective application...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0000000000000137

    authors: Endes S,Bachler M,Li Y,Mayer C,Hanssen H,Hametner B,Schmidt-Trucksäss A,Wassertheurer S

    更新日期:2015-10-01 00:00:00

  • Blood pressure cuff comparability study.

    abstract:BACKGROUND:Manufacturer-supplied blood pressure (BP) cuffs are part of the automatic oscillometric BP devices algorithm. MATERIALS AND METHODS:This study assessed the differences in BP values using the Omron HEM 907-XL (Omron) device with two types of cuffs: the Baum cuff (BC) and the supplied Omron cuff (OC). A sampl...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章,随机对照试验

    doi:10.1097/MBP.0000000000000208

    authors: Ostchega Y,Nwankwo T,Zhang G,Chiappa M

    更新日期:2016-12-01 00:00:00

  • Frequency domain of heart rate and blood pressure variability in essential hypertensive patients during sleep: differences between dippers and non-dippers.

    abstract::METHODS: Autonomic nervous function was evaluated by means of power spectral analysis of heart rate and blood pressure variability in dipper (n = 10) and non-dipper (n = 9) essential hypertensive subjects during sleep. The non-dipper subjects were defined as those in whom the nocturnal decrease in blood pressure was <...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Frisina N,Pedullà M,Mento G,Lanuzza B

    更新日期:1996-10-01 00:00:00

  • Propaq Neonatal monitor used with its own single- or with Critikon twin-hose cuffs: does it matter?.

    abstract::Some oscillometric non-invasive blood pressure (NIBP) monitors use twin-hose cuffs, whereas others use single-hose cuffs. The incompatibility hinders the transfer of patients from monitors using a single-hose cuff to monitors using twin-hose cuffs and vice versa. In particular, the Propaq Encore uses a single-hose cuf...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Amoore JN

    更新日期:1997-12-01 00:00:00

  • Methods of data analysis of ambulatory blood pressure monitoring in clinical trials.

    abstract::The objectives of the analysis are to compare the blood pressure levels at different phases of the trial and to be sure that the effect will continue until the next dose. There are problems before the analysis for which consensus is necessary. If the overall 24 h study is the first generally recognized stage, it is di...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: de Gaudemaris R,Chau NP

    更新日期:1996-06-01 00:00:00

  • Morning hypertension in chronic kidney disease is sustained type, but not surge type.

    abstract:OBJECTIVE:We have shown that as renal function deteriorates, the circadian blood pressure (BP) rhythm shifts to a nondipper pattern and the duration until nocturnal BP decline [dipping time (DT)] is prolonged. We investigated whether or not morning hypertension (BP 2 h after awakening >135/85 mmHg) in chronic kidney di...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/MBP.0b013e32835026c6

    authors: Mizuno M,Fukuda M,Miura T,Wakamatsu T,Naito T,Sato R,Togawa H,Sasakawa Y,Tomonari T,Ono M,Kato Y,Ichikawa T,Shirasawa Y,Ito A,Yoshida A,Kimura G

    更新日期:2012-02-01 00:00:00

  • Clinical test of a blood pressure measurement device with a wrist cuff.

    abstract::OBJECTIVE: To test a semi-automatic oscillometric blood pressure measurement device with a wrist cuff. METHODS: Two different test methods, partly based on European and American standards for clinical investigation of non-invasive sphygmomanometers, were worked out and compared. Mostly German guidelines were considere...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:

    authors: Wonka F,Thümmler M,Schöppe A

    更新日期:1996-08-01 00:00:00

  • Validation of the Omron HEM-907 device for blood pressure measurement.

    abstract:BACKGROUND:The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. DESIGN:The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, ...

    journal_title:Blood pressure monitoring

    pub_type: 杂志文章

    doi:10.1097/00126097-200208000-00006

    authors: El Assaad MA,Topouchian JA,Darné BM,Asmar RG

    更新日期:2002-08-01 00:00:00

  • Serum copeptin levels in predicting nondippers in newly diagnosed hypertension.

    abstract:OBJECTIVES:The nondipper pattern in hypertension is associated with worse cardiovascular outcomes. In this study, we hypothesized that high copeptin levels could predict nondippers among hypertensive patients and investigated the associations between copeptin levels and nondipper pattern in newly diagnosed hypertensive...

    journal_title:Blood pressure monitoring

    pub_type: 临床试验,杂志文章

    doi:10.1097/MBP.0000000000000123

    authors: Uzun F,Biyik I,Akturk IF,Yalcin AA,Erturk M,Oner E,Kalkan AK,Yalcin B,Atmaca H

    更新日期:2015-08-01 00:00:00