Atypical "tails-up" capnograph due to breach in the sampling tube of side-stream capnometer.

Abstract:

OBJECTIVE:An atypical "tails-up" capnograph pattern was noticed in a patient during the use of an accidentally crushed sampling tube with a slit-like hole. We investigated the mechanics involved in the observed capnograph pattern. METHODS:Forty consenting ASA I patients of both sexes presenting for tonsillectomy were included in this study. After intravenous induction of anaesthesia, intermittent positive pressure ventilation (IPPV) using a mechanical ventilator was maintained for 20 min and the capnograph trace, ETCO2, and inspiratory/expiratory sevoflurane were compared using a breached sampling tube and then an intact sampling tube. Similar comparisons were made during spontaneous breathing. RESULTS:During IPPV, an atypical "tails-up" capnograph was noted using the breached sampling tubing. At similar inspiratory sevoflurane (2.0 +/- 0.03) levels, expiratory levels (0.9 +/- 0.03) were significantly lower when using the breached sampling tube than the intact tube (1.7 +/- 0.03). ETCO2 with the breached sampling tube (26.8 +/- 0.30 mmHg) showed significantly lower values than with the intact sampling tube (37 +/- 0.3). During spontaneous breathing, the capnograph was normal in shape with both sampling tubes, but ETCO2 and both insp./exp. sevoflurane levels were lower with the breached sampling tube. CONCLUSION:During IPPV, pressure in the breathing circuit is lower during exhalation, thus allowing air to enter through the slit-like hole in the sampling tube causing erroneously low ETCO2 and expiratory sevoflurane. With inspiration, positive pressure in the breathing circuit, transmitted to the sampling tube, prevents air admixture and the upsurge in CO2 is displayed giving the capnograph an atypical "tails-up" appearance. During spontaneous breathing, since pressure in the breathing circuit barely becomes positive during exhalation and is negative during inspiration, air mixes with the sampled gas during both phases and so the capnograph shape was normal but with lower values for ETCO2, insp./exp. sevoflurane, and nitrous oxide levels. If undiagnosed, this defect in the sampling tube can lead to significant errors in the measurement of inspired and expired gas concentrations.

journal_name

J Clin Monit Comput

authors

Tripathi M,Pandey M

doi

10.1023/a:1009990210114

subject

Has Abstract

pub_date

2000-01-01 00:00:00

pages

17-20

issue

1

eissn

1387-1307

issn

1573-2614

journal_volume

16

pub_type

杂志文章
  • Leakage of sevoflurane from vaporizer detected by air monitoring system: a case report.

    abstract::Leakage of inhalational anesthetic gases is an important issue not only to staff health but awareness during general anesthesia. After inhalation of sevoflurane commenced, the audible and visible alarm of our custom-made air monitoring system was activated. The system measured 2-4 ppm of sevoflurane. A thorough search...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-013-9479-9

    authors: Tanaka M,Yamaura K,Higashi M,Hoka S

    更新日期:2013-12-01 00:00:00

  • Characterization of respiratory compromise and the potential clinical utility of capnography in the post-anesthesia care unit: a blinded observational trial.

    abstract::The utility of capnography to detect early respiratory compromise in surgical patients after anesthesia is unclear due to limited prospective data. The purpose of this trial was to determine the frequency and duration of capnography-detected respiratory adverse events in the post-anesthesia care unit (PACU). In this p...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-019-00333-9

    authors: Chung F,Wong J,Mestek ML,Niebel KH,Lichtenthal P

    更新日期:2020-06-01 00:00:00

  • Perioperative echocardiography-guided hemodynamic therapy in high-risk patients: a practical expert approach of hemodynamically focused echocardiography.

    abstract::The number of high-risk patients undergoing surgery is growing. To maintain adequate hemodynamic functioning as well as oxygen delivery to the vital organs (DO2) amongst this patient population, a rapid assessment of cardiac functioning is essential for the anesthesiologist. Pinpointing any underlying cardiovascular p...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章,评审

    doi:10.1007/s10877-020-00534-7

    authors: Trauzeddel RF,Ertmer M,Nordine M,Groesdonk HV,Michels G,Pfister R,Reuter D,Scheeren TWL,Berger C,Treskatsch S

    更新日期:2020-05-26 00:00:00

  • Using physiological models and decision theory for selecting appropriate ventilator settings.

    abstract:OBJECTIVE:To present a decision support system for optimising mechanical ventilation in patients residing in the intensive care unit. METHODS:Mathematical models of oxygen transport, carbon dioxide transport and lung mechanics are combined with penalty functions describing clinical preference toward the goals and side...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-006-9049-5

    authors: Rees SE,Allerød C,Murley D,Zhao Y,Smith BW,Kjaergaard S,Thorgaard P,Andreassen S

    更新日期:2006-12-01 00:00:00

  • Eye lens dosimetry in anesthesiology: a prospective study.

    abstract::The eye lens is one of the most sensitive organs for radiation injury and exposure might lead to radiation induced cataract. Eye lens dosimetry in anesthesiology has been published in few clinical trials and an active debate about the causality of radiation induced cataract is still ongoing. Recently, the Internationa...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-016-9857-1

    authors: Vaes B,Van Keer K,Struelens L,Schoonjans W,Nijs I,Vandevenne J,Van Poucke S

    更新日期:2017-04-01 00:00:00

  • Performance of a capnodynamic method estimating effective pulmonary blood flow during transient and sustained hypercapnia.

    abstract::The capnodynamic method is a minimally invasive method continuously calculating effective pulmonary blood flow (COEPBF), equivalent to cardiac output when intra pulmonary shunt flow is low. The capnodynamic equation joined with a ventilator pattern containing cyclic reoccurring expiratory holds, provides breath to bre...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-017-0021-3

    authors: Sigmundsson TS,Öhman T,Hallbäck M,Redondo E,Sipmann FS,Wallin M,Oldner A,Hällsjö Sander C,Björne H

    更新日期:2018-04-01 00:00:00

  • Investigation of optical neuro-monitoring technique for detection of maintenance and emergence states during general anesthesia.

    abstract::The American Society of Anesthesiologist recommends peripheral physiological monitoring during general anesthesia, which offers no information regarding the effects of anesthetics on the brain. Since no "gold standard" method exists for this evaluation, such a technique is needed to ensure patient comfort, procedure q...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-017-9998-x

    authors: Hernandez-Meza G,Izzetoglu M,Osbakken M,Green M,Abubakar H,Izzetoglu K

    更新日期:2018-02-01 00:00:00

  • Computerised respiratory sounds can differentiate smokers and non-smokers.

    abstract::Cigarette smoking is often associated with the development of several respiratory diseases however, if diagnosed early, the changes in the lung tissue caused by smoking may be reversible. Computerised respiratory sounds have shown to be sensitive to detect changes within the lung tissue before any other measure, howev...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-016-9887-8

    authors: Oliveira A,Sen I,Kahya YP,Afreixo V,Marques A

    更新日期:2017-06-01 00:00:00

  • Protection of the remaining spinal cord function with intraoperative neurophysiological monitoring during paraparetic scoliosis surgery: a case report.

    abstract:OBJECTIVE:To demonstrate the usefulness of rectus femoris muscle MEPs monitoring in a paraparetic neuromuscular scoliosis case. METHODS:Multiple monitoring modalities including SEPs, MEPs and EMG were performed for an anterior and posterior correction surgery for a neuromuscular scoliosis patient with no motor and sen...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-011-9325-x

    authors: Chen Z,Lerman J

    更新日期:2012-02-01 00:00:00

  • Effects of intraoperative positive end-expiratory pressure optimization on respiratory mechanics and the inflammatory response: a randomized controlled trial.

    abstract::Applying lung protective mechanical ventilation (LPV) during general anaesthesia even in patients with non-injured lungs is recommended. However, the effects of an individual PEEP-optimisation on respiratory mechanics, oxygenation and their potential correlation with the inflammatory response and postoperative complic...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-020-00519-6

    authors: Ruszkai Z,Kiss E,László I,Bokrétás GP,Vizserálek D,Vámossy I,Surány E,Buzogány I,Bajory Z,Molnár Z

    更新日期:2020-05-09 00:00:00

  • Comparison of Univent tube and EZ blocker in one lung ventilation; airway pressures and gas exchange.

    abstract::Univent tube (UT) and EZ-blocker were used for one-lung ventilation (OLV). UT is a single lumen tube with a small separate lumen containing a bronchial blocker. EZ-blocker differs with its unique y-shaped double-cuffed distal end. We aimed to compare these two airway devices effects on airway pressures, oxygenation, v...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章,随机对照试验

    doi:10.1007/s10877-017-0028-9

    authors: Hoşten T,Aksu C,Kuş A,Cesur S,Türkyılmaz N,Solak M

    更新日期:2018-04-01 00:00:00

  • Pulmonary air embolism.

    abstract::Pulmonary air embolism is a well-known consequence of surgery, trauma, diving, and aviation. This article reviews the physiological effects, means of detection and methods of prevention and treatment of pulmonary air embolism. The primary physiological effects are elevated pulmonary artery pressures, increased ventila...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章,评审

    doi:10.1023/a:1011455701892

    authors: Souders JE

    更新日期:2000-01-01 00:00:00

  • Monitoring the microcirculation.

    abstract::One of the main goals of hemodynamic support is to preserve tissue perfusion. However issue perfusion is related more to microvascular perfusion than aortic blood flow. Monitoring the microcirculation has long been difficult. Recent technologic advances have made feasible monitoring of the microcirculation at bedside ...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章,评审

    doi:10.1007/s10877-012-9383-8

    authors: De Backer D,Donadello K,Cortes DO

    更新日期:2012-10-01 00:00:00

  • Teaching respiratory physiology: clinical correlation with a human patient simulator.

    abstract::In recent years students have increasingly objected to laboratory exercises involving animal subjects. We have replaced the valuable animal experiments with demonstrations using a full-scale human patient simulator. In small groups first-year medical students observe realistic clinical situations such as opioid-induce...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1023/a:1011401011887

    authors: Euliano TY

    更新日期:2000-01-01 00:00:00

  • Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique.

    abstract:OBJECTIVES:Bioreactance, the analysis of intrabeat variations in phase of a transthoracic voltage in response to an applied high frequency transthoracic current, was recently introduced for noninvasive cardiac output measurement (NICOM). We evaluated NICOM compared to thermo- dilution (TD) in several clinical settings....

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章,多中心研究

    doi:10.1007/s10877-008-9112-5

    authors: Raval NY,Squara P,Cleman M,Yalamanchili K,Winklmaier M,Burkhoff D

    更新日期:2008-04-01 00:00:00

  • Trans-ocular brain impedance index for assessment of cerebral autoregulation in a porcine model of cerebral hemodynamic perturbation.

    abstract::Cerebrovascular autoregulation (CA) is often impaired following traumatic brain injury. Established technologies and metrics used to assess CA are invasive and conducive for measurement, but not for continuous monitoring. We developed a trans-ocular brain impedance (TOBI) method that may provide non-invasive and conti...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-020-00556-1

    authors: Tiba MH,McCracken BM,Leander DC,Colmenero CI,Cummings BC,Greer NL,Picton P,Williamson CA,Ward KR

    更新日期:2020-07-14 00:00:00

  • Multi-parameter vital sign database to assist in alarm optimization for general care units.

    abstract::Continual vital sign assessment on the general care, medical-surgical floor is expected to provide early indication of patient deterioration and increase the effectiveness of rapid response teams. However, there is concern that continual, multi-parameter vital sign monitoring will produce alarm fatigue. The objective ...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-015-9790-8

    authors: Welch J,Kanter B,Skora B,McCombie S,Henry I,McCombie D,Kennedy R,Soller B

    更新日期:2016-12-01 00:00:00

  • Sympatho-vagal balance, as quantified by ANSindex, predicts post spinal hypotension and vasopressor requirement in parturients undergoing lower segmental cesarean section: a single blinded prospective observational study.

    abstract::Hypotension subsequent to spinal anesthesia occurs in a significant number of parturients undergoing lower segment caesarian section. Currently available methods to predict the incidence of hypotension, its severity and the outcome are sub-optimal. Many workers have used basal heart rate as one of the predictors. But ...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-016-9906-9

    authors: Prashanth A,Chakravarthy M,George A,Mayur R,Hosur R,Pargaonkar S

    更新日期:2017-08-01 00:00:00

  • Effective hypertensive treatment using data mining in Saudi Arabia.

    abstract::In the present investigation, the data sets of NCD (Non Communicable Diseases) risk factors, a standard report of Saudi Arabia 2005, in collaboration with WHO (World Health Organisation) were employed. The Oracle Data Miner (ODM) tool was used for the analysis and prediction of data. The data sets for different age gr...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-010-9260-2

    authors: Almazyad AS,Ahamad MG,Siddiqui MK,Almazyad AS

    更新日期:2010-12-01 00:00:00

  • Validation of a mathematical model for understanding intracranial pressure curve morphology.

    abstract::The physiology underlying the intracranial pressure (ICP) curve morphology is not fully understood. Recent research has suggested that the morphology could be dependent on arterial cerebral inflow and the physiological and pathophysiological properties of the intracranial cavity. If understood, the ICP curve could pro...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-019-00342-8

    authors: Unnerbäck M,Ottesen JT,Reinstrup P

    更新日期:2020-06-01 00:00:00

  • Electrical impedance tomography's correlation to lung volume is not influenced by anthropometric parameters.

    abstract:STUDY OBJECTIVES:Electrical impedance tomography (EIT) is able to reflect physiological parameters such as real-time changes in global and regional lung volume. EIT can aid in the assessment of lung recruitment, and its use has been validated in preliminary studies monitoring mechanical ventilation at the bedside. ICU ...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-006-9021-4

    authors: Marquis F,Coulombe N,Costa R,Gagnon H,Guardo R,Skrobik Y

    更新日期:2006-06-01 00:00:00

  • The pressure exerted on the tongue during intubation with simultaneous cervical spine immobilisation: a comparison between four videolaryngoscopes and the Macintosh laryngoscope-a manikin study.

    abstract::Excessive pressures exerted on the tongue during intubation may be associated with serious complications and may make intubation more difficult. The aim of this study was to evaluate five different intubating devices in the hands of experienced anaesthetists during simulated conditions of reduced cervical spine mobili...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-017-0095-y

    authors: Aleksandrowicz D,Gaszyński T

    更新日期:2018-10-01 00:00:00

  • Two-site regional oxygen saturation and capnography monitoring during resuscitation after cardiac arrest in a swine pediatric ventricular fibrillatory arrest model.

    abstract:ABSTARCT:To investigate the use of two-site regional oxygen saturations (rSO2) and end tidal carbon dioxide (EtCO2) to assess the effectiveness of resuscitation and return of spontaneous circulation (ROSC). Eight mechanically ventilated juvenile swine underwent 28 ventricular fibrillatory arrests with open cardiac mass...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-019-00291-2

    authors: Al-Subu AM,Hacker TA,Eickhoff JC,Ofori-Amanfo G,Eldridge MW

    更新日期:2020-02-01 00:00:00

  • Effects of prolonged ischemia on human skeletal muscle microcirculation as assessed by near-infrared spectroscopy.

    abstract::Near-infrared spectroscopy (NIRS) has been used to detect in vivo microvascular alterations by means of a vascular occlusion test. We sought to analyse by NIRS the microcirculatory profile of patients undergoing prolonged tourniquet-induced bloodless condition for extremity surgery, and compare the results with time o...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-016-9877-x

    authors: Tujjar O,De Gaudio AR,Tofani L,Di Filippo A

    更新日期:2017-06-01 00:00:00

  • Bi-spectral index guided closed-loop anaesthesia delivery system (CLADS) in pheochromocytoma.

    abstract:OBJECTIVES:Anaesthetic management of pheochromocytoma is far from satisfactory even though various techniques are available. Feasibility as well as safety of Bi-spectral Index (BIS)-guided closed-loop control of anaesthesia has been shown in various clinical settings. We evaluated the performance of BIS-guided closed-l...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-009-9181-0

    authors: Hegde HV,Puri GD,Kumar B,Behera A

    更新日期:2009-08-01 00:00:00

  • Focus on renal blood flow in mechanically ventilated patients with SARS-CoV-2: a prospective pilot study.

    abstract::Mechanically ventilated patients with ARDS due to the severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) seem particularly susceptible to AKI. Our hypothesis was that the renal blood flow could be more compromised in SARS-CoV-2 patients than in patients with "classical" ARDS. We compared the renal resistivit...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-020-00633-5

    authors: Fogagnolo A,Grasso S,Dres M,Gesualdo L,Murgolo F,Morelli E,Ottaviani I,Marangoni E,Volta CA,Spadaro S

    更新日期:2021-01-01 00:00:00

  • Impact of non invasive and beat-to-beat arterial pressure monitoring on intraoperative hemodynamic management.

    abstract::The Nexfin device allows for non-invasive beat-to-beat blood pressure monitoring (BP(NXF)). Perioperative hypotension and hypertension have been shown to be associated with poor clinical outcomes. The goal of the present study was to assess the ability of this device to decrease the duration of significant intraoperat...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-012-9344-2

    authors: Chen G,Chung E,Meng L,Alexander B,Vu T,Rinehart J,Cannesson M

    更新日期:2012-04-01 00:00:00

  • Transoesophageal Doppler monitoring for fluid and hemodynamic treatment during lung surgery.

    abstract:INTRODUCTION:Patients undergoing lung resection are vulnerable to fluid overhydration. Recently, goal-directed fluid therapy using transoesophageal Doppler monitoring (TDM) has been shown to improve postoperative clinical outcome. The aim of this study was to assess the feasibility of TDM during open-chest procedures f...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-008-9144-x

    authors: Diaper J,Ellenberger C,Villiger Y,Robert J,Tschopp JM,Licker M

    更新日期:2008-10-01 00:00:00

  • Comparison of the use of AnaConDa® versus AnaConDa-S® during the post-operative period of cardiac surgery under standard conditions of practice.

    abstract::Changes have been made to the AnaConDa device (Sedana Medical, Stockholm, Sweden), decreasing its size to reduce dead space and carbon dioxide (CO2) retention. However, this also involves a decrease in the surface area of the activated carbon filter. The CO2 elimination and sevoflurane (SEV) reflection of the old devi...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-019-00285-0

    authors: Marcos-Vidal JM,Merino M,González R,García C,Rey S,Pérez I

    更新日期:2020-02-01 00:00:00

  • A motor evoked potential trending system may discriminate outcome: retrospective application with three cases.

    abstract::This report presents a method for tracking Motor Evoked Potential (MEP) amplitudes over the course of a case using a moving least squares linear regression (LSMAs). During a case, newly obtained MEP amplitudes are compared to those predicted by a just previous linear regression (least squares moving average or LSMA). ...

    journal_title:Journal of clinical monitoring and computing

    pub_type: 杂志文章

    doi:10.1007/s10877-018-0181-9

    authors: Jasiukaitis P,Lyon R

    更新日期:2019-06-01 00:00:00