Clinical performance evaluation of the new hematology analyzer Mindray BC-6000.

Abstract:

INTRODUCTION:With the progression of blood analysis technology, hematology analyzers become more complex and diverse. How to choose a superb instrument is a challenge for the laboratories. In the essay, we studied whether the newest BC-6000 hematology analyzer meets the needs of a clinical hematology laboratory. METHODS:Methods comparison was performed using 350 blood samples from patients between different measurement procedures; the basic analytical performance was also tested, including the throughput, carryover, precision, and linearity in different modes. The flagging performances for blasts, immature granulocytes, and NRBC were compared with manual microscopy. RESULTS:There were minimal carryover (<0.30%) and excellent actual blood linearity for all routinely used parameters concerned by the clinicians (R2  ≥ 0.997). Repeatability and reproducibility were satisfactory at all testing levels. The functional sensitivity of leukocytes and platelets in the blood and leukocytes and erythrocytes in body fluid was excellent at the 20% CV level. BC-6000 and XN displayed very high correlations for complete blood count (CBC) parameters and very high consistency for leukocyte differentials and NRBC compared with manual microscopy. BC-6000 showed excellent sensitivity and specificity flagging ability on blasts (82.9% and 82.4%) and NRBC (80.0% and 96.9%). For immature granulocytes, BC-6000 showed excellent sensitivity but common specificity flagging ability (91.7% and 65.6%). CONCLUSION:The clinical performance of BC-6000 is excellent, and the analyzer can provide timely and accurate reporting for most of the small- to large-scale laboratories.

journal_name

Int J Lab Hematol

authors

Shen Y,Cao J,Zhou Z,Wang Y,Shen Y,He J

doi

10.1111/ijlh.13075

subject

Has Abstract

pub_date

2019-10-01 00:00:00

pages

622-634

issue

5

eissn

1751-5521

issn

1751-553X

journal_volume

41

pub_type

杂志文章
  • The clinical utility of new reticulocyte and erythrocyte parameters on the Sysmex XN 9000 for iron deficiency in pregnant patients.

    abstract:INTRODUCTION:Iron deficiency (ID) is a major cause of morbidity in pregnancy. Antenatal clinics use a hemoglobin to screen for ID, which delays the diagnosis of subclinical ID. The aim of this study was to investigate the clinical utility of the percentage of microcytic red cells (%Micro-R), percentage of hypochromic r...

    journal_title:International journal of laboratory hematology

    pub_type: 临床试验,杂志文章

    doi:10.1111/ijlh.12904

    authors: Levy S,Schapkaitz E

    更新日期:2018-12-01 00:00:00

  • Inappropriate use of protein C, protein S, and antithrombin testing for hereditary thrombophilia screening: an experience from a large university hospital.

    abstract:INTRODUCTION:Deficiencies of protein C, protein S, and antithrombin are the main inherited risk factors in Thai patients with venous thromboembolism, although the prevalence is not high. METHODS:To evaluate the appropriate use of the testing for these proteins, the test orders of 503 patients were retrospectively revi...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2011.01332.x

    authors: Tientadakul P,Chinthammitr Y,Sanpakit K,Wongwanit C,Nilanont Y

    更新日期:2011-12-01 00:00:00

  • Transillumination: a new tool to eliminate the impact of venous stasis during the procedure for the collection of diagnostic blood specimens for routine haematological testing.

    abstract:INTRODUCTION:The collection of diagnostic blood specimens for routine haematological testing (RHT) is traditionally performed with tourniquet. However, the transillumination devices based on cold near-infrared LEDs have been formerly proposed as a valuable tool for identifying reliable venous accesses, especially in pa...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2011.01305.x

    authors: Lima-Oliveira G,Lippi G,Salvagno GL,Montagnana M,Scartezini M,Guidi GC,Picheth G

    更新日期:2011-10-01 00:00:00

  • Optimal wavelength for the clot waveform analysis: Determination of the best resolution with minimal interference of the reagents.

    abstract:INTRODUCTION:Clot waveform analysis (CWA), a new methodology to assess coagulation process, can be usefully applied in various clinical settings. However, its clinical use is limited mainly because of the absence of standardization. No consensus exists regarding the wavelengths at which CWA has to be performed what is ...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12975

    authors: Evrard J,Siriez R,Morimont L,Thémans P,Laloy J,Bouvy C,Gheldof D,Mullier F,Dogné JM,Douxfils J

    更新日期:2019-06-01 00:00:00

  • The HemoScreen hematology point-of-care device is suitable for rapid evaluation of acute leukemia patients.

    abstract:BACKGROUND:Hematological patients, receiving intensive chemotherapy (predominantly acute leukemia patients), have repeated postchemotherapy periods with severe bone marrow suppression. As a result, these patients require regular monitoring of the complete blood counts (CBC) for optimal patient care. To reduce the strai...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13330

    authors: Kristian Kur D,Thøgersen D,Kjeldsen L,Friis-Hansen L

    更新日期:2021-02-01 00:00:00

  • Factor XIII-A subunit Val34Leu polymorphism in fatal hemorrhagic stroke.

    abstract::Blood coagulation factor XIII (FXIII) plays a key role in the protection of fibrin clot against fibrinolysis, in the cross-linking of fibrin and its mechanical strength. The role of the FXIII-A subunit Val34Leu polymorphism with fatal primary intracerebral hemorrhages (PICH) has not been studied. We evaluated retrospe...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01465.x

    authors: Antalfi B,Pongrácz E,Csiki Z,Mezei ZA,Shemirani AH

    更新日期:2013-02-01 00:00:00

  • Nontransferrin-bound iron in transfused patients with sickle cell disease.

    abstract::The value of nontransferrin-bound iron (NTBI) as an index of iron overload in patients with thalassemia has been evaluated; however, data in patients with sickle cell disease (SCD) is limited. NTBI levels were evaluated in a cross-sectional study of 43 transfused patients with SCD. Patient charts were reviewed for dem...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2010.01224.x

    authors: Inati A,Musallam KM,Cappellini MD,Duca L,Taher AT

    更新日期:2011-04-01 00:00:00

  • The biological characteristics of dendritic cells derived in vitro from myelogeneous leukemia cells and healthy donor cells.

    abstract::Successful adoptive immunotherapy for leukemia depends on the generation of T cells that can specifically react with malignant cells. Dendritic cells (DCs) are important antigen-presenting cells in the development of antileukemia T-cell responses. Mononuclear cells (MNC) were isolated from peripheral blood or bone mar...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.00986.x

    authors: Tong XM,Yao HP,Qian WB,Zhu LF,Fu ZH,Huang ZL,Jin J

    更新日期:2008-10-01 00:00:00

  • Three novel alternative splicing mutations in BCR-ABL1 detected in CML patients with resistance to kinase inhibitors.

    abstract:INTRODUCTION:Multiple types of mutations in the BCR-ABL1 kinase domain have been reported. We previously reported a common alternatively spliced BCR-ABL mRNA with a 35-nucleotide insertion (35INS). We report three novel alternative splicing mutants expressed as the dominant transcripts in patient with chronic myelogeno...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2010.01291.x

    authors: Ma W,Giles F,Zhang X,Wang X,Zhang Z,Lee TS,Yeh CH,Albitar M

    更新日期:2011-06-01 00:00:00

  • International Council for Standardization in Haematology (ICSH) laboratory guidance for the evaluation of haemostasis analyser-reagent test systems. Part 1: Instrument-specific issues and commonly used coagulation screening tests.

    abstract::Before a new method is used for clinical testing, it is essential that it is evaluated for suitability for its intended purpose. This document gives guidance for the performance of verification, validation and implementation processes required by regulatory and accreditation bodies. It covers the planning and executio...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.13411

    authors: Gardiner C,Coleman R,de Maat MPM,Dorgalaleh A,Echenagucia M,Gosselin RC,Ieko M,Kitchen S

    更新日期:2020-11-29 00:00:00

  • Toward routine detection of extracellular vesicles in clinical samples.

    abstract::The majority, if not all, of human cell types secrete extracellular vesicles (EVs) into their environment, at least partly as a means of intercellular communication. These secreted vesicles can be detected in most bodily fluids including blood, urine, and saliva. The number of secreted vesicles and their composition i...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.12247

    authors: van der Meel R,Krawczyk-Durka M,van Solinge WW,Schiffelers RM

    更新日期:2014-06-01 00:00:00

  • A hierarchical approach in the diagnostic workflow of chronic myelomonocytic leukemia: Pivotal role of the "Mono-dysplasia-score" combined with flow cytometric quantification of monocyte subsets.

    abstract:INTRODUCTION:Monocytosis is a frequent trigger for blood smear review in a routine hematology laboratory whereas chronic myelomonocytic leukemia (CMML) is infrequent and arises mostly in elderly patients. In order to define the best workflow for monocytosis, we studied three diagnostic approaches: the classical morphol...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13115

    authors: Zhu J,Sourdeau E,Aubert H,Clauser S,Maillon A,Capron C,Jondeau K,Ronez E,Schillinger F,Garnache-Ottou F,Cornet E,Bardet V

    更新日期:2019-12-01 00:00:00

  • A level-headed approach to measuring direct oral anticoagulants: A 2-year retrospective analysis of DOAC levels from a tertiary UK centre.

    abstract:INTRODUCTION:Direct oral anticoagulants (DOAC) are commonly prescribed and measuring drug levels may be useful in a number of contexts. However, data on DOAC level measurement and their clinical utility in real-world studies are limited. METHODS:We carried out a 2-year retrospective cohort study of DOAC levels measure...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12944

    authors: Denny NDR,Keighley L,Siganporia Z,Thachil J,Nash MJ

    更新日期:2019-04-01 00:00:00

  • Analysis of polymorphisms of TNF-α, LT-α, IL-10, IL-12 and CTLA-4 in patients with warm autoimmune haemolytic anaemia.

    abstract:INTRODUCTION:Autoimmune haemolytic anaemia (AIHA) is defined as the increased destruction of red blood cells (RBCs) in the presence of anti-RBC autoantibodies and/or complement. Its pathogenesis is multifactorial and includes changes in mechanisms of cytokine production and functionality. A number of recent studies hav...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01400.x

    authors: D'Abronzo LS,Barros MM,Bordin JO,Figueiredo MS

    更新日期:2012-08-01 00:00:00

  • Paediatric reference intervals for common coagulation assays in Chinese children as performed on the STA-R coagulation analyzer.

    abstract:INTRODUCTION:In order to correctly manage the paediatric patients affected with haemostatic disorders, age-appropriate reference intervals should be used. The purpose of this study was to establish age-specific reference intervals for prothrombin time (PT), thrombin time (TT), activated partial thromboplastin time (aPT...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13098

    authors: Liu J,Dai Y,Yuan E,Li Y,Wang Q,Wang L,Su Y

    更新日期:2019-10-01 00:00:00

  • The mutual control of iron and erythropoiesis.

    abstract:BACKGROUND:Iron is essential for hemoglobin synthesis during terminal erythropoiesis. To supply adequate iron the carrier transferrin is required together with transferrin receptor endosomal cycle and normal mitochondrial iron utilization. Iron and iron protein deficiencies result in different types of anemia. Iron-def...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.12505

    authors: Camaschella C,Pagani A,Nai A,Silvestri L

    更新日期:2016-05-01 00:00:00

  • Entonox as a sedative for bone marrow aspiration and biopsy.

    abstract::Bone marrow aspiration and biopsy can be a painful procedure. Sedation techniques may make this investigation more acceptable to patients, but have the potential to cause life-threatening complications, as well as requiring additional staff and equipment for safe administration. We assessed the use of Entonox, a 50 : ...

    journal_title:International journal of laboratory hematology

    pub_type: 临床试验,杂志文章

    doi:10.1111/j.1751-553X.2007.00905.x

    authors: Gudgin EJ,Besser MW,Craig JI

    更新日期:2008-02-01 00:00:00

  • Direct international normalized ratio determination using multicalibrators is more responsive than the conventional method for measuring prothrombin time.

    abstract::Direct international normalized ratio (INR) determination using certified INR plasmas was shown to improve precision and accuracy. We evaluated the utility of a multicalibrator in determining INR. INR values were measured in 493 blood samples from patients subjected to anticoagulation therapy (320) and control subject...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2009.01195.x

    authors: Cha CH,Park CJ,Kim DH,Kim MJ,Cho YU,Jang S,Chi HS

    更新日期:2010-08-01 00:00:00

  • Abnormal scattergrams and cell population data generated by fully automated hematological analyzers: New tools for screening malaria infection?

    abstract:INTRODUCTION:Malaria is a life-threatening infectious disease, which has been for long confined to specific endemic areas. Nevertheless, the recent increase in immigration flows from endemic regions and imported cases has reemphasized many diagnostic challenges in Western countries, thus paving the way to introduce rap...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12790

    authors: Buoro S,Manenti B,Seghezzi M,Moioli V,Bagorria M,Callegaro A,Ottomano C,Lippi G

    更新日期:2018-06-01 00:00:00

  • Analytical comparison between two hematological analyzer systems: CAL-8000 vs. XN-9000.

    abstract:INTRODUCTION:This study was aimed to compare the analytical performance of traditional and new parameters and morphological flags of CAL-8000 and XN-9000. The automated differential leukocyte count (DIFF profile) and morphological flags were compared with optical microscopy (OM). METHODS:A total of 1025 peripheral blo...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12597

    authors: Buoro S,Mecca T,Seghezzi M,Manenti B,Azzarà G,Dominoni P,Crippa A,Ottomano C,Lippi G

    更新日期:2017-04-01 00:00:00

  • Falciparum malaria parasitemia index for predicting severe malaria.

    abstract:INTRODUCTION:While hyperparasitemia is considered an important indicator for the development of severe malaria, there is currently no consensus on the quantitative definition of hyperparasitemia. This study was conducted to establish a cutoff point for peripheral parasitemia among patients with Plasmodium falciparum ma...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2011.01398.x

    authors: Tangpukdee N,Krudsood S,Kano S,Wilairatana P

    更新日期:2012-06-01 00:00:00

  • Guidance for storing blood samples in laboratories performing complete blood count with differential.

    abstract:INTRODUCTION:The complete blood count (CBC) with differential leukocyte count (DIFF) is an important part of clinical laboratory analyses and provides crucial data for clinicians. Delivery time after blood collection and conditions of storage is known to affect the reliability of results of some hematologic parameters....

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01452.x

    authors: Cornet E,Behier C,Troussard X

    更新日期:2012-12-01 00:00:00

  • Paroxysmal nocturnal haemoglobinuria. Experience over a 10 years period.

    abstract:INTRODUCTION:Paroxysmal nocturnal hemoglobinuria (PNH) is a hemolytic, clonal and acquired disorder of the hematopoietic stem cell with a deficiency of all glycophosphatidyl-inositol (GPI) linked proteins. The aim of this retrospective study was to analyse haematological and biochemical data from 152 patients referred ...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12156

    authors: Canalejo K,Riera Cervantes N,Felippo M,Sarandría C,Aixalá M

    更新日期:2014-04-01 00:00:00

  • Use of a capillary specimen in the laboratory to verify a point-of-care international normalized ratio: Avoidance of a venipuncture in a pediatric setting.

    abstract:INTRODUCTION:Point-of-care (POC) international normalized ratio (INR) values above an institutional cutoff are confirmed in the laboratory using a gold standard venous specimen. This can be problematic in a pediatric setting. METHOD:In this study, 449 consecutive POC INR results were compared to an INR performed in th...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12920

    authors: Williams VK

    更新日期:2019-02-01 00:00:00

  • Comparison of five point-of-care D-dimer assays with the standard laboratory method.

    abstract:INTRODUCTION:D-dimer (DD) assays are effective for the exclusion of deep-vein thrombosis (DVT), but point-of-care (POC) DD assays have not been fully evaluated. METHODS:We have compared five POC DD assays (Pathfast, Cardiac, Vidas, Stratus and NycoCard) with our routine DD method (Tinaquant), testing 60 samples from p...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01421.x

    authors: Antovic JP,Höög Hammarström K,Forslund G,Eintrei J,Sten-Linder M

    更新日期:2012-10-01 00:00:00

  • Novel chromosomal translocations in multiple myeloma: t(13;16)(q14;q24) and t(1;15)(q10;q26).

    abstract::Multiple myeloma (MM) is a malignant plasma cell disorder that involves multiple genetic abnormalities. Chimeric transcription factors, created by gene fusion as a result of chromosomal translocations, have been implicated in the pathogenesis of the disease. Here, we report the conventional cytogenetic analysis of a M...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2006.00837.x

    authors: Akalin I,Kutlay NY,Ilhan O,Tukun A

    更新日期:2007-06-01 00:00:00

  • The presence of leukaemia-associated phenotypes is an independent predictor of induction failure in acute myeloid leukaemia.

    abstract::Immunophenotyping of acute myeloid leukaemia (AML) has controversial implications with regards to prognosis. The aims of the present study were to determine the frequency of leukaemia-associated phenotypes (LAP) in AML and to correlate their presence with response to induction chemotherapy. We analysed bone marrow sam...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.01003.x

    authors: Al-Mawali A,To LB,Gillis D,Hissaria P,Mundy J,Lewis I

    更新日期:2009-02-01 00:00:00

  • Red cell membrane disorders.

    abstract::Significant advances have been made in our understanding of the structural basis for altered cell function in various inherited red cell membrane disorders with reduced red cell survival and resulting hemolytic anemia. The current review summarizes these advances as they relate to defining the molecular and structural...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.12657

    authors: Narla J,Mohandas N

    更新日期:2017-05-01 00:00:00

  • Accuracy of serum transferrin receptor levels in the diagnosis of iron deficiency among hospital patients in a population with a high prevalence of thalassaemia trait.

    abstract::The diagnosis of iron deficiency in hospital patients can be difficult in the presence of inflammation. A raised serum transferrin receptor (sTfR) level is useful as a marker of iron deficiency as it is unaffected by inflammation. However, diseases that cause an increase in erythropoietic activity can also result in a...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.00982.x

    authors: Ong KH,Tan HL,Tam LP,Hawkins RC,Kuperan P

    更新日期:2008-12-01 00:00:00

  • Technological advances in diagnostic testing for von Willebrand disease: new approaches and challenges.

    abstract::Diagnostic tests for von Willebrand disease (VWD) are important for the assessment of VWD, which is a commonly encountered bleeding disorder worldwide. Technical innovations have been applied to improve the precision and lower limit of detection of von Willebrand factor (VWF) assays, including the ristocetin cofactor ...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.12220

    authors: Hayward CP,Moffat KA,Graf L

    更新日期:2014-06-01 00:00:00