Evaluating the analytical performance of five new coagulation assays for the measurement of prothrombin time and activated thromboplastin time.

Abstract:

INTRODUCTION:New methods for coagulation tests require careful assessment before routine use. We evaluated the analytical performance of five new coagulation assays for measuring prothrombin time (PT) and activated partial thromboplastin time (aPTT). METHODS:PT Rec, PT Owren, aPTT, aPTT Lupus and aPTT Screen assays (Roche Diagnostics) were evaluated on cobas t 711 and cobas t 511 analysers (Roche Diagnostics) at four European centres. Analytical performance and method comparisons with relevant commercially available assays were performed to Clinical Laboratory Standards Institute guidelines using residual anonymized samples. Lot-to-lot comparison and equivalency of the cobas t analysers were also assessed; reference ranges were determined using samples from apparently healthy volunteers. RESULTS:Overall, coefficients of variation were ≤1.3% for within-run precision and ≤6.3% for total reproducibility across all sites. Deming regression analyses showed good agreement between each assay (cobas t 711) and respective comparator method (Pearson's r: 0.964-0.999, n > 120 samples/assay/site). Passing-Bablok regression analyses demonstrated equivalence of the two cobas t platforms for use with each assay (Pearson's r ≥ 0.995). Lot-to-lot consistency was high for all assays and comparisons (Pearson's r ≥ 0.998). Reference ranges (2.5th-97.5th percentiles; n = 200 samples/assay) in seconds were 8.4-10.6 (PT Rec), 18.2-27.2 (PT Owren), 23.6-30.6 (aPTT), 24.1-31.7 (aPTT Lupus) and 23.9-33.2 (aPTT Screen). CONCLUSION:Based on the excellent analytical performance and good agreement with relevant comparator methods, the five coagulation assays on the novel cobas t 711 and cobas t 511 analysers are suitable for routine use in core laboratories.

journal_name

Int J Lab Hematol

authors

Kitchen S,Geisen U,Kappelmayer J,Quehenberger P,Drieß J,Lowe A,Jones R,Boehm JG,Miles G,Rozsnyai G

doi

10.1111/ijlh.12897

subject

Has Abstract

pub_date

2018-12-01 00:00:00

pages

645-654

issue

6

eissn

1751-5521

issn

1751-553X

journal_volume

40

pub_type

杂志文章
  • The impact of long-haul air travel on variables of the athlete's biological passport.

    abstract:INTRODUCTION:Dehydration, fluid shifts or changes in coagulation occurring during air travel can trigger distinct reactions in the haematological system. Athletes are concerned that these effects might impair sporting performance, increase the risk of thrombosis or cause abnormalities in blood values that might be mist...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01450.x

    authors: Schumacher YO,Klodt F,Nonis D,Pottgiesser T,Alsayrafi M,Bourdon PC,Voss SC

    更新日期:2012-12-01 00:00:00

  • Upregulation of long noncoding RNA Lnc-IRF2-3 and Lnc-ZNF667-AS1 is associated with poor survival in B-chronic lymphocytic leukemia.

    abstract:BACKGROUND:Lnc-IRF2-3 and Lnc-ZNF667-AS1 were recently studied as a positive biomarker for many tumor cells. However, experimental studies found that they are associated with worse outcomes in B-CLL. METHODS:A prospective case study was conducted on 135 B-CLL patients that were compared to thirty healthy controls. The...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13167

    authors: El-Khazragy N,Esmaiel MA,Mohamed MM,Hassan NS

    更新日期:2020-06-01 00:00:00

  • The biological characteristics of dendritic cells derived in vitro from myelogeneous leukemia cells and healthy donor cells.

    abstract::Successful adoptive immunotherapy for leukemia depends on the generation of T cells that can specifically react with malignant cells. Dendritic cells (DCs) are important antigen-presenting cells in the development of antileukemia T-cell responses. Mononuclear cells (MNC) were isolated from peripheral blood or bone mar...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.00986.x

    authors: Tong XM,Yao HP,Qian WB,Zhu LF,Fu ZH,Huang ZL,Jin J

    更新日期:2008-10-01 00:00:00

  • How I investigate for bleeding disorders.

    abstract:INTRODUCTION:Laboratory investigations for bleeding disorders are warranted when an individual has a personal and/or family history of bleeding, and/or laboratory findings that suggest the possibility of an inherited or acquired bleeding disorder. METHODS:This review summarizes author's experience with ordering and re...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.12822

    authors: Hayward CPM

    更新日期:2018-05-01 00:00:00

  • Assessing maternal clotting function with novel global coagulation assays: A prospective pilot study.

    abstract:INTRODUCTION:Women are at higher risk of venous thromboembolism (VTE) during pregnancy and the puerperium. Global coagulation assays (GCAs), including thromboelastography (TEG), thrombin generation using the calibrated automated thrombogram (CAT) and fibrin generation using the overall haemostatic potential assay (OHP)...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13389

    authors: O'Keefe D,Lim HY,Tham J,Ho P,Hui L

    更新日期:2020-11-10 00:00:00

  • Evaluation of the leukocyte differential on a new automated flow cytometry hematology analyzer.

    abstract:INTRODUCTION:The Hematoflow (Beckman Coulter, USA) is a new automated hematology analyzer, which provides a 16-part white blood cell count (WBC) differential. METHODS:We evaluated the differential WBC count performance of the Hematoflow. 101 blood samples from patients were selected for comparison analysis. RESULTS:T...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01432.x

    authors: Kim JE,Kim BR,Woo KS,Han JY

    更新日期:2012-10-01 00:00:00

  • Evaluation of a new venom-based clotting assay of protein C.

    abstract::Congenital protein C deficiency significantly increases the risk of venous thromboembolism, a serious and potentially lethal condition. Protein C levels can be determined by chromogenic, clotting and antigenic assays, each type of assay has differences in specificity and sensitivity to protein C deficiency. In princip...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.00972.x

    authors: Cooper PC,Cooper SM,Goodfellow KJ,Hickey KP,Kitchen S,Makris M

    更新日期:2008-10-01 00:00:00

  • A level-headed approach to measuring direct oral anticoagulants: A 2-year retrospective analysis of DOAC levels from a tertiary UK centre.

    abstract:INTRODUCTION:Direct oral anticoagulants (DOAC) are commonly prescribed and measuring drug levels may be useful in a number of contexts. However, data on DOAC level measurement and their clinical utility in real-world studies are limited. METHODS:We carried out a 2-year retrospective cohort study of DOAC levels measure...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12944

    authors: Denny NDR,Keighley L,Siganporia Z,Thachil J,Nash MJ

    更新日期:2019-04-01 00:00:00

  • The automation of routine light transmission platelet aggregation.

    abstract:INTRODUCTION:The investigation of platelet function by aggregometry requires specialist equipment and is labour intensive. We have developed an automated platelet aggregation method on a routine coagulation analyser. METHODS:We used a CS-2000i (Sysmex) with prototype software to perform aggregation in platelet-rich pl...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12161

    authors: Lawrie AS,Kobayashi K,Lane PJ,Mackie IJ,Machin SJ

    更新日期:2014-08-01 00:00:00

  • Comparison of five point-of-care D-dimer assays with the standard laboratory method.

    abstract:INTRODUCTION:D-dimer (DD) assays are effective for the exclusion of deep-vein thrombosis (DVT), but point-of-care (POC) DD assays have not been fully evaluated. METHODS:We have compared five POC DD assays (Pathfast, Cardiac, Vidas, Stratus and NycoCard) with our routine DD method (Tinaquant), testing 60 samples from p...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2012.01421.x

    authors: Antovic JP,Höög Hammarström K,Forslund G,Eintrei J,Sten-Linder M

    更新日期:2012-10-01 00:00:00

  • Establishment of reference intervals for pediatric complete capillary blood counts: A multicenter study in Beijing.

    abstract:INTRODUCTION:Clinical reference intervals represent the normal range of clinical parameters for distinguishing healthy and sick individuals, and they show some variation among different populations. Many reference intervals are still lacking for the pediatric population in China. Thus, the aim of this study was to esta...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13150

    authors: Sun Q,Yue Y,Tian Y,Wang Q

    更新日期:2020-10-01 00:00:00

  • Proteasome inhibitor bortezomib overcomes P-gp-mediated multidrug resistance in resistant leukemic cell lines.

    abstract:INTRODUCTION:To study the effect of bortezomib alone or in combination with daunorubicin (DNR) on an mdr1 single-factor drug-resistant leukemia cell line K562/MDR1, a multifactor-resistant cell line K562/A02, a drug-sensitive cell line K562, and primary cells from acute myeloid leukemia patients. METHODS:The cell line...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2011.01384.x

    authors: Zheng B,Zhou R,Gong Y,Yang X,Shan Q

    更新日期:2012-06-01 00:00:00

  • Performance evaluation of the new hematology analyzer Sysmex XN-series.

    abstract:INTRODUCTION:The Sysmex XN-series is a new automated hematology analyzer designed to improve the accuracy of cell counts and the specificity of the flagging events. METHODS:The basic characteristics and the performance of new measurement channels of the XN were evaluated and compared with the Sysmex XE-2100 and the ma...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12254

    authors: Seo JY,Lee ST,Kim SH

    更新日期:2015-04-01 00:00:00

  • Optimal wavelength for the clot waveform analysis: Determination of the best resolution with minimal interference of the reagents.

    abstract:INTRODUCTION:Clot waveform analysis (CWA), a new methodology to assess coagulation process, can be usefully applied in various clinical settings. However, its clinical use is limited mainly because of the absence of standardization. No consensus exists regarding the wavelengths at which CWA has to be performed what is ...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12975

    authors: Evrard J,Siriez R,Morimont L,Thémans P,Laloy J,Bouvy C,Gheldof D,Mullier F,Dogné JM,Douxfils J

    更新日期:2019-06-01 00:00:00

  • Limitation of the activated partial thromboplastin time as a monitoring method of the direct thrombin inhibitor argatroban.

    abstract:BACKGROUND:Argatroban is licensed for patients with heparin-induced thrombocytopenia and monitoring is conventionally by activated partial thromboplastin time (APTT) ratio with a target of 1.5-3.0 and not exceeding 100 s. The APTT may be influenced by coagulopathies, lupus anticoagulant and raised FVIII levels. Variabl...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12414

    authors: Guy S,Kitchen S,Maclean R,Van Veen JJ

    更新日期:2015-12-01 00:00:00

  • Paediatric reference intervals for common coagulation assays in Chinese children as performed on the STA-R coagulation analyzer.

    abstract:INTRODUCTION:In order to correctly manage the paediatric patients affected with haemostatic disorders, age-appropriate reference intervals should be used. The purpose of this study was to establish age-specific reference intervals for prothrombin time (PT), thrombin time (TT), activated partial thromboplastin time (aPT...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13098

    authors: Liu J,Dai Y,Yuan E,Li Y,Wang Q,Wang L,Su Y

    更新日期:2019-10-01 00:00:00

  • Clinical and hematological characteristics of 88 patients with COVID-19.

    abstract:INTRODUCTION:To retrospectively analyze epidemiological, clinical and hematological characteristics of COVID-19 patients. METHODS:The demographic, symptoms, and physiological parameters of 88 patients were collected and analyzed. The performance of complete blood count (CBC) indexes for monitoring and predicting the s...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13291

    authors: Zhang H,Cao X,Kong M,Mao X,Huang L,He P,Pan S,Li J,Lu Z

    更新日期:2020-12-01 00:00:00

  • Direct international normalized ratio determination using multicalibrators is more responsive than the conventional method for measuring prothrombin time.

    abstract::Direct international normalized ratio (INR) determination using certified INR plasmas was shown to improve precision and accuracy. We evaluated the utility of a multicalibrator in determining INR. INR values were measured in 493 blood samples from patients subjected to anticoagulation therapy (320) and control subject...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2009.01195.x

    authors: Cha CH,Park CJ,Kim DH,Kim MJ,Cho YU,Jang S,Chi HS

    更新日期:2010-08-01 00:00:00

  • The role of automated measurement of red cell subpopulations on the Sysmex XE 5000 analyzer in the differential diagnosis of microcytic anemia.

    abstract::Cell counter-based formulae have been used in the differential diagnosis of microcytic anemia. The measurement of new red cell parameters is now available on the Sysmex XE 5000 analyzer. Derived from the percentages of microcytic and hypochromic red cells, the authors describe the new formula %microcytic-%hypochromic,...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2010.01237.x

    authors: Urrechaga E,Borque L,Escanero JF

    更新日期:2011-02-01 00:00:00

  • Flow cytometry data analysis: Recent tools and algorithms.

    abstract::Flow cytometry (FCM) allows scientists to rapidly quantify up to 50 parameters for millions of cells per sample. The bottleneck in the application of the technology is data analysis, and the high number of parameters measured by the current generation of instruments requires the use of advanced computational algorithm...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.13016

    authors: Montante S,Brinkman RR

    更新日期:2019-05-01 00:00:00

  • Abnormal scattergrams and cell population data generated by fully automated hematological analyzers: New tools for screening malaria infection?

    abstract:INTRODUCTION:Malaria is a life-threatening infectious disease, which has been for long confined to specific endemic areas. Nevertheless, the recent increase in immigration flows from endemic regions and imported cases has reemphasized many diagnostic challenges in Western countries, thus paving the way to introduce rap...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12790

    authors: Buoro S,Manenti B,Seghezzi M,Moioli V,Bagorria M,Callegaro A,Ottomano C,Lippi G

    更新日期:2018-06-01 00:00:00

  • ADAMTS13 activity to von Willebrand factor antigen ratio predicts acute kidney injury in patients with COVID-19: Evidence of SARS-CoV-2 induced secondary thrombotic microangiopathy.

    abstract:INTRODUCTION:Severe COVID-19 is often compounded by a prothrombotic state that is associated with poor outcomes. In this investigation, we aimed to evaluate ADAMTS13 activity, von Willebrand factor level (VWF:Ag), and the corresponding ADAMTS13 activity/VWF:Ag ratio, in patients with COVID-19 and for associations with ...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13415

    authors: Henry BM,Benoit SW,de Oliveira MHS,Lippi G,Favaloro EJ,Benoit JL

    更新日期:2020-12-03 00:00:00

  • Sickle cell-haemoglobin E (HbSE) compound heterozygosity: a clinical and haematological study.

    abstract::The paucity of clinical reports in the world literature suggests that, as a disease entity, haemoglobin SE compound heterozygosity is of negligible importance. In view of the significant community prevalence of this haemoglobinopathy in the Sultanate of Oman where it is the second most prevalent sickling disorder, a h...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/j.1365-2257.2006.00886.x

    authors: Knox-Macaulay HH,Ahmed MM,Gravell D,Al-Kindi S,Ganesh A

    更新日期:2007-08-01 00:00:00

  • Assessing nonvitamin K antagonist oral anticoagulants (NOACs) in the laboratory.

    abstract::Nonvitamin K antagonist oral anticoagulants (NOACs) are being used with increasing frequency due to their safety profile, ease of use, and given that therapeutic monitoring is not required. As these agents have only recently been FDA approved, their effect on routine and specialty coagulation assays is not well apprec...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章,评审

    doi:10.1111/ijlh.12350

    authors: Gosselin RC,Adcock DM

    更新日期:2015-05-01 00:00:00

  • Different kinetic patterns of BCR-ABL transcript levels in imatinib-treated chronic myeloid leukemia patients after achieving complete cytogenetic response.

    abstract::Bone marrow BCR-ABL transcript levels were monitored serially by real-time quantitative PCR in 46 imatinib-treated chronic myeloid leukemia patients after achieving complete cytogenetic response (CCyR) for a median of 42 months (range: 9-53). Of 41 patients in continuous CCyR, 32 and nine could achieve a >/=3-log (MMo...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.00962.x

    authors: Qin YZ,Liu YR,Zhu HH,Li JL,Ruan GR,Zhang Y,Jiang Q,Jiang H,Li LD,Chang Y,Huang XJ,Chen SS

    更新日期:2008-08-01 00:00:00

  • Use of Sysmex XN-10 red blood cell parameters for screening of hereditary red blood cell diseases and iron deficiency anaemia.

    abstract:INTRODUCTION:In daily practice in haematology laboratories, red blood cell (RBC) abnormalities are frequent and their management is a real challenge. The aim of this study is to establish a "decision tree" using RBC and reticulocyte parameters from the SYSMEX XN-10 analyser to distinguish between patients with a heredi...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.13278

    authors: Nivaggioni V,Bouriche L,Coito S,Le Floch AS,Ibrahim-Kosta M,Leonnet C,Arnoux I,Loosveld M

    更新日期:2020-12-01 00:00:00

  • The importance of β globin deletion analysis in the evaluation of patients with β thalassemia.

    abstract:INTRODUCTION:Beta globin deletion/duplication analysis may serve as a useful adjunct to sequence analysis. Our purpose was to develop a robust assay for beta globin deletion/duplication analysis and determine its role in evaluating patients with beta thalassemia. METHODS:A single tube semi-quantitative fluorescent PCR...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2010.01287.x

    authors: Mikula M,Buller-Burckle A,Gallivan M,Sun W,Franklin CR,Strom CM

    更新日期:2011-06-01 00:00:00

  • Paroxysmal nocturnal haemoglobinuria. Experience over a 10 years period.

    abstract:INTRODUCTION:Paroxysmal nocturnal hemoglobinuria (PNH) is a hemolytic, clonal and acquired disorder of the hematopoietic stem cell with a deficiency of all glycophosphatidyl-inositol (GPI) linked proteins. The aim of this retrospective study was to analyse haematological and biochemical data from 152 patients referred ...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12156

    authors: Canalejo K,Riera Cervantes N,Felippo M,Sarandría C,Aixalá M

    更新日期:2014-04-01 00:00:00

  • Expression of CD4 is correlated with an unfavorable prognosis in wild-type NPM1, FLT3-ITD-negative cytogenetically normal adult acute myeloid leukemia.

    abstract:INTRODUCTION:In the cytogenetically normal population of AML (CN-AML), FLT3-ITD-positive and wild-type NPM1 is correlated with a worse outcome, and FLT3-ITD-negative with NPM1-mut is correlated with a better outcome. This leaves a large subpopulation of CN-AML patients without NPM1 or FLT3-ITD mutations with heterogene...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/ijlh.12649

    authors: Guo RJ,Atenafu EG,Schimmer AD,Minden MD,Chang H

    更新日期:2017-08-01 00:00:00

  • Platelet size has diagnostic predictive value for bone marrow metastasis in patients with solid tumors.

    abstract::Though not very common, solid tumor involvement of the bone marrow (BM) may have serious consequences. Recent studies have shown that mean platelet volume (MPV) is a good indicator for BM disease in the differential diagnosis of thrombocytopenia. We investigated the significance of MPV in the diagnosis of BM metastasi...

    journal_title:International journal of laboratory hematology

    pub_type: 杂志文章

    doi:10.1111/j.1751-553X.2007.00947.x

    authors: Aksoy S,Kilickap S,Hayran M,Harputluoglu H,Koca E,Dede DS,Erman M,Turker A

    更新日期:2008-06-01 00:00:00